中文摘要: 目前国内主流电子测血压技术多采用下气测量,和上气测量技术相比,测量时间长且加压的最大值高,使被测者不适感增强。因此研究上气测血压原理,详细介绍其算法实现,深入分析其PWM 加压、butterworth 滤波等关键技术。最后研究证明,通过该技术实现的血压测量系统测量时间短,功耗低,被测者不适感减少,测量效果完全满足设计需求。
中文关键词: 血压 上气测量 测量时间 PWM 加压 butterworth 滤波
Abstract:At present, most of the Medical device companies measure blood pressure during deflation process, which occupates more measuring time, higher pressure, less comfortable, compared to inflation measurement. So we study inflation measurement theory, detail the algorithm, analyze the key technology of PWM pressure method and butterworth filter. It is proved to be less time-consuming, lower power consumption and more comfortable.
keywords: blood pressure inflation measurement measurement time PWM pressure filter based on butterworth