DOI:
计算机系统应用英文版:2012,21(2):196-199
本文二维码信息
码上扫一扫!
基于示波法测量血压的算法改进
(1.中国科学院研究生院,北京 100049;2.中国科学院沈阳计算技术研究所,沈阳 110168)
Algorithm Improvement of Blood Pressure Measurement Based on Oscillography
(1.Graduate University, Chinese Academy of Sciences, Beijing 100049, China;2.Shenyang Institute of Computing Technology, Chinese Academy of Sciences, Shenyang 110168, China)
摘要
图/表
参考文献
相似文献
本文已被:浏览 2087次   下载 7260
Received:May 24, 2011    Revised:June 15, 2011
中文摘要: 示波法血压测量的关键是脉搏波的提取以及收缩压、舒张压的判定。就此两方面,改进并实现了非线性的形态滤波器,从脉搏波和袖带压的混合信号中滤得脉搏波,并结合波形特征法和S 判别法,实现了收缩压舒张压的判定。通过验证,较单纯的归一化比例系数法,该算法明显提高了测量结果的准确性与重复性。
Abstract:The crucial factors of blood pressure measurement based on oscillography are the extraction of pulse waves and the determination algorithm. So this paper improved the nonlinear filter based on morphology, to extract pulse waves successfully. We combined wave characteristics and sphygmo-oscillographic method, realizing the determination algorithm of blood pressure. Experiments prove this algorithm effectively improved the measurement accuracy and repeatability than normalization coefficient method.
文章编号:     中图分类号:    文献标志码:
基金项目:
引用文本:
王维维,蒲宝明,李生金,贺宝岳.基于示波法测量血压的算法改进.计算机系统应用,2012,21(2):196-199
WANG Wei-Wei,PU Bao-Ming,LI Sheng-Jin,HE Bao-Yue.Algorithm Improvement of Blood Pressure Measurement Based on Oscillography.COMPUTER SYSTEMS APPLICATIONS,2012,21(2):196-199